Research Peptides UK: when laboratories measure the same sample
Shared samples, from C-peptide in serum to a reference antibody, reveal what calibration, reproducibility and agreement between laboratories actually describe.
When the same sample goes to several laboratories, results spread. In shared-sample studies, common calibration and reference materials narrowed the spread but did not remove it: three laboratories still differed by a median 13 per cent for C-peptide and 22 per cent for insulin. One laboratory’s figure, without a stated uncertainty, is a single draw from that distribution.
A shared sample puts the measuring system itself on trial
Send one material to many laboratories and the question changes. Instead of asking only what a laboratory found, you can ask how far the answers spread, which methods travel well between sites, and where disagreement begins. A result can be beautifully repeatable inside one building and still awkward to reproduce in the next.
A 2015 study did this on a grand scale. Eleven laboratories, running fourteen liquid-chromatography and mass-spectrometry systems, developed and tested assays for 125 peptides representing 27 cancer-relevant proteins in human plasma. Median within-laboratory and between-laboratory reproducibility was below 20 per cent.1 The study also shows how much happens before the instrument: median peptide loss was about 70 per cent during digestion, the enzymatic step that cuts proteins into peptides, and about 27 per cent during desalting.1 A peptide signal was being used to stand in for a protein quantity, so every loss on the way mattered.
A common reference can expose a calibration difference
C-peptide is a more focused case. It is released alongside insulin, so clinicians and researchers use it to study how much insulin the pancreas itself is making. Different laboratories can report it in the same unit and still reach different numbers. Calibration is the relationship that turns an instrument’s signal into a quantity, and that is where the trouble lies.
A 2017 review traced efforts to build a reference system for C-peptide: pure primary material, reference measurement procedures, and serum-based secondary materials linking them to routine results. The UK’s National Institute for Biological Standards and Control prepared the first WHO International Standard for human C-peptide, with 8.64 micrograms per ampoule and a stated uncertainty, made to calibrate immunoassays rather than to act as a primary standard. By July 2014 the standardisation committee had shown that serum-based secondary reference materials could reduce variability among assay methods.2
The serum matters. A purified molecule and the same molecule surrounded by blood proteins are different analytical materials, and a comparison material must behave enough like real samples to connect results usefully.
Modified calibrator molecules changed what was being counted
A 2022 study made one calibration difference concrete. Two certified C-peptide reference materials contained mostly unmodified C-peptide plus smaller amounts of modified forms, 1.9 and 5.6 per cent respectively. Their certified values counted the related forms together; the mass-spectrometry assay counted only the unmodified molecule.3
In 45 patient serum samples, calibrating on the total quantity produced a noticeable bias compared with calibrating on unmodified C-peptide; switching to the unmodified figure reduced it whichever manufacturer’s material was used. A comparison of 39 patient samples with the University of Missouri laboratory then showed good agreement, with a slope of 0.94.3 The modified forms include pyroglutamylated C-peptide, which loses a molecule of water and so changes mass and behaviour on the column.
If two numbers count different molecules, they can disagree before a single unknown sample is measured.On Wu et al., 2022
This is the measurement-science version of a lesson made elsewhere on this site: a label or a certified value describes a defined quantity, and it pays to know exactly which one. The pillar guide’s section on purity versus content makes the same point about powders.
Between-laboratory spread remained after harmonisation
A 2024 study compared antibody-free mass-spectrometry measurements of C-peptide and insulin across three laboratories, using shared materials and leftover clinical samples with a common calibration.4 Imprecision means scatter: here, how far the laboratories’ results spread around each other.
Three laboratories, two molecules
Median between-laboratory imprecision (coefficient of variation) after common calibration, from Moradian et al., 2024. Bar length is proportional to the percentage.
Scale runs from 0 to 25 per cent. Averaging repeat measurements narrowed the spread; it did not remove it, and the difference between the two molecules stayed visible.
A shared calibrator supplied a common starting point. Averaging repeats reduced the scatter. Neither made the laboratories identical, and insulin, the larger molecule, remained the less reproducible of the two.
One antibody, seventy-six laboratories, many views
The NISTmAb is a monoclonal antibody that the US National Institute of Standards and Technology distributes as a reference material. In an interlaboratory study of its glycosylation, the sugar structures attached to the protein, 76 laboratories submitted 103 reports using methods of their own choosing. Laboratories identified between 4 and 48 glycan compositions each; across the exercise 116 compositions were reported, and consensus values could be assigned to 57.5
That spread is not a scandal. It is what happens when competent groups look at a complex molecule through different instruments, and it is why consensus values are built from many laboratories rather than borrowed from one.
Repeatable is not the same as reproducible
A 2019 comparison studied the antigen-binding fragment of the same antibody by hydrogen–deuterium exchange mass spectrometry, a technique that follows how quickly hydrogens in a protein swap with deuterium from heavy water, which reports on its shape and flexibility. Fifteen laboratories took part. Together they reported 430 peptide sequences covering the whole fragment, yet nearly half were reported by only one laboratory and just two by every laboratory.6
Within laboratories, repeatability was tight: most achieved mass repeatability of about 0.15 dalton. Across all fifteen, reproducibility of the corrected deuterium uptake was about 9 per cent, and 6.5 per cent for a nine-laboratory group working at the same temperature.6 Repeatability describes clustering under the same conditions; reproducibility asks what survives a change of laboratory. The two are reported separately for a reason.
Three questions to put to any single number
These studies measured different properties of different materials, but they converge on three questions worth asking of any analytical result, including the one-line figures on a research certificate:
| Material | Property | Comparison question |
|---|---|---|
| Human plasma (Abbatiello 2015) | 125 peptides from 27 proteins | Can a multiplexed assay transfer between 11 laboratories? |
| C-peptide reference materials (Wu 2022) | Modified vs unmodified forms | Which molecule does the certified value count? |
| Human serum (Moradian 2024) | C-peptide and insulin | How closely do three laboratories agree after common calibration? |
| NISTmAb (De Leoz 2019) | Glycan compositions | What do different methods report for one molecule? |
| NISTmAb Fab (Hudgens 2019) | Deuterium uptake | How reproducible is a structural signal between sites? |
- What exactly is being counted? The intended molecule only, or related forms too?
- Against what reference? A traceable standard, a consensus value, or the laboratory’s own material?
- With what spread? Is there an uncertainty, and would a second competent laboratory land inside it?
A certificate that answers none of these is a single draw from a distribution nobody has described. For how that applies to peptide powders, see the pillar guide; for why this title names no sellers, see our page on sellers; for the vocabulary, the glossary. Our sister title peptidesjournal.uk covers the wider history of peptide chemistry.
Questions readers ask
What is the difference between repeatability and reproducibility?
Repeatability describes how tightly results cluster when the same laboratory repeats a measurement under the same conditions. Reproducibility asks how well results agree when conditions change, for example when a different laboratory, instrument or operator measures the same sample.
Why does a shared calibrator not make laboratories agree perfectly?
Because calibration fixes only the starting relationship between signal and quantity. Sample preparation, instrument behaviour and the exact molecular form being counted still differ between sites, so some scatter remains even after harmonisation.
What does this mean for a certificate from a single laboratory?
That one laboratory’s number is a single draw from a distribution. Without a stated uncertainty or a comparison with other laboratories, a reader cannot tell how far another competent laboratory’s answer might sit from it.
References
- Abbatiello SE, Schilling B, Mani DR, Zimmerman LJ, Hall SC, MacLean B, et al. Large-scale interlaboratory study to develop, analytically validate and apply highly multiplexed, quantitative peptide assays to measure cancer-relevant proteins in plasma. Mol Cell Proteomics. 2015;14(9):2357–2374. DOI
- Little RR, Wielgosz RI, Josephs R, Kinumi T, Takatsu A, Li H, Stein D, Burns C. Implementing a reference measurement system for C-peptide: successes and lessons learned. Clin Chem. 2017;63(9):1447–1456. DOI
- Wu Z, Kabytaev K, Mu J, Connolly S, Clarke NJ, Little R, McPhaul MJ. Critical need to assess modified and un-modified peptides in C-peptide standard materials. J Mass Spectrom Adv Clin Lab. 2022;26:7–8. DOI
- Moradian A, Goonatileke E, Lin T, Hatten-Beck M, Emrick M, Schepmoes AA, Fillmore TL, et al. Interlaboratory comparison of antibody-free LC-MS/MS measurements of C-peptide and insulin. Clin Chem. 2024;70(6):855–864. DOI
- De Leoz MLA, Duewer DL, et al. NIST interlaboratory study on glycosylation analysis of monoclonal antibodies: comparison of results from diverse analytical methods. Mol Cell Proteomics. Published online 7 October 2019. DOI
- Hudgens JW, Karageorgos IL, Gallagher ES, Anderson KW, Filliben JJ, et al. Interlaboratory comparison of hydrogen–deuterium exchange mass spectrometry measurements of the Fab fragment of NISTmAb. Anal Chem. 2019;91(11):7336–7345. DOI